Anderson Manufacturing AV-9000 Recording Equipment User Manual


 
Edition 3
Rev G
Section 11
11-7
CHART Will place a dot on the inner or outer ring, whichever is selected for calibration
SCROLL Will select either the inner or outer ring
UP Will move the selected (inner or outer) ring position to the left and print a dot in the new
position without moving the chart.
DOWN Will move the selected (inner or outer) ring position to the right and print a dot in the new
position without moving the chart.
RESET Will move the chart 30 steps while not printing any dots
ESC Will exit with no change to the calibration values
ENTER Will exit with the new values being stored
Step 3
With INNER RING CALIBRATE on the upper display line, depress the to view the current location of the
inner ring zero point. Use UP/DOWN to position the dot accordingly, use RESET to move the chart.
Step 4
Depress the to toggle the upper display line to read OUTER RING CALIBRATE.
Step 5
Depress the to view the current location of the outer ring zero point. Use UP/DOWN to position the dot
accordingly, use RESET to move the chart.
Step 6
Depress the to save calibration values.
11.5 MILLIAMP OUTPUT CALIBRATION
This routine is used to adjust the mA Output(s) for span. A milliamp meter is required to perform this calibration.
Step 1
With mA OUTPUTS in the lower display line, depress the and CAL WHICH mA OUTPUT
appears in the upper display line and 1 appears in the lower display line. To change,
depress the , then the to desired output number, then depress .