Agilent Technologies L4400 Guitar User Manual


 
L4433A Dual/Quad 4x8 Reed Matrix 5
L4400 User’s Guide 119
One-Wire Mode
To physically configure the module in 1- wire mode, use the 34933T- 002
terminal block, or a compatible standard or custom cable. If using a
standard or custom cable, make sure you connect interlock pins 17 and
33 on the Matrix 2 D- sub connector. Refer to the pinout drawing and
table on page 125.
In 1-wire mode, the L4433A module contains four matrices (1 through 4),
each with 32 1- wire crosspoint non- latching reed relays organized in a 4- row
by 8- column configuration. Every row and column has one wire each. Each
crosspoint relay has a unique channel number representing the matrix, and
the single- wire row and column that intersect to make the crosspoint. For
example, channel 218 represents Matrix 2, row 1 and column 8. See the
simplified schematic on page 124.
In 1- wire mode, you can close no more than 40 channels simultaneously due
to power dissipation. For example, with one analog bus relay closed you can
close up to a maximum of 39 channel relays. If you try to close more than the
allowed number of channels, you will receive an error message.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 3 and Matrix 4 in 1- wire mode of this module connect
to the analog buses. By closing channels 921 and 922 you can connect row 1
and row 2 respectively to the HI (H) and LO (L) lines of ABus1 and ABus2.
You can connect multiple matrix modules externally and/or through the
analog buses for applications that require large matrices. For information on
linking multiple matrix modules, refer to page 116 of this chapter.
When the power is off, matrix relays and analog bus relays open.