Agilent Technologies L4400 Guitar User Manual


 
76 L4400 User’s Guide
3 Operating and Programming
You can store at least 500,000 readings in memory and all readings are
automatically time stamped. If memory overflows, the new readings will
overwrite the first (oldest) readings stored; the most recent readings are
always preserved.
You can specify a trigger count in conjunction with a sweep count. The two
parameters operate independent of one another, and the total number of
readings returned will be the product of the two parameters.
The CONFigure and MEASure? commands automatically set the scan
trigger count to 1.
The instrument sets the scan trigger count to 1 after a Factory Reset (*RST
command). An Instrument Preset (SYSTem:PRESet command) or Card
Reset (SYSTem:CPON command) does not change the setting.
The following command sets the trigger count:
TRIGger:COUNt
To configure a continuous scan, send TRIG:COUNT INFINITY.
Sweep Count
Usage:
L4450A 64- Bit Digital I/O (digital input, counter channels only)
L4452A Multifunction Module (digital input, totalizer channels only)
The sweep count sets the number of sweeps per trigger event during a scan (a
sweep is one pass through the scan list).
Figure 3-11. Sweep Count Diagram.
The sweep count is valid only while scanning. If no channels have been
assigned to the scan list, the specified sweep count is ignored (no error is
generated).
t
Sweep 1Trigger
Sweep Count
(1 to 500,000 sweeps)
Sweep 2
Sweep n
Trigger
. . .