44 L4400 User’s Guide
2 Software Installation and Configuration
Figure 2-24. Instrument Firmware Update Complete.
5. Once the carrier update AND instrument sub-assembly update (if one
occurs) complete and no instrument activity is indicated by the front panel
LEDs, cycle power on the instrument. Once the power- on sequence completes,
select ‘Refresh All’ in the Agilent Connection Expert (Figure 2- 10).
Agilent Connection Expert may report that the instrument’s configuration has
changed. This is represented by a yellow triangle and an exclamation point (!)
next to the updated instrument. Select (highlight) the instrument name. Select
‘Change Properties...’ and then click either ‘Test Connection’ or ‘Identify
Instrument’ to update Agilent Connection Expert and then click ‘OK’. Repeat
for each updated instrument on the LAN and GPIB interfaces.