Konica Minolta CM-2600d Drums User Manual


 
E-15
Preparation
Storing a Target Mask
In the case of the CM-2600d, a Target Mask that is not in use can be attached to the Target Mask mount
section of the White Calibration Plate, so that it can be stored together with the instrument.
M
emo
Even in the case of the CM-2500d, when the Target Mask is removed for cleaning the integrating sphere, it can be
attached to the
Target Mask mount section of the White Calibration Plate to prevent loss and damage.
Note
Do not touch the inner surface of the Target Mask or allow it to get dirty and scratched.
Do not exert excessive force on the latch of the Target Mask. Doing so may damage the latch, disabling use of
the Target Mask.
Target Mask
Latch
Target Mask mount section
White
Calibration
Plate
Notch
Attaching a Target Mask
1.
2
11
Place the Target Mask on the Target Mask mount
section so that the inner surface of the latch is
aligned with the notch on the White Calibration
Plate.
2. Hold the outer edge of the mask and turn it clock-
wise to secure it.
Removing the Target Mask
1.
1
1
1
2
2
2
Hold the outer edge of the mask and turn it coun-
ter-clockwise until the inner surface of the latch is
aligned with the notch on the White Calibration
Plate.
2. Hold the outer edge of the mask and remove it.