Konica Minolta CM-2600d Drums User Manual


 
E-42
Setting a Measurement Condition
Setting the Delay Time
12
Turn B to set the desired delay time, then press A.
<Settings>
• DELAY TIME:
The delay time is used to prevent influences caused by movement of the hands, and is the duration from
when the MEAS. button is pressed until the lamp flashes and measurement is performed. Specify the de
-
sired delay time (0.1 to 3.0 seconds in steps of 0.1 seconds). If “0.0” (second) is specified, no delay time
will be provided.
Keeping B held down to left or right will cause the value to change continuously.
“BREAK” will be highlighted.
13
<Performing Calibration>
Turn B to select “PREV”, then press A.
Setting will be completed and the
<COND
*
> screen will appear.
“T*” is the number of the color difference target data for the next measurement. If nothing is selected,
“T---” appears.