Konica Minolta CM-2600d Drums User Manual


 
E-86
Assessments
5
If they are satisfactory, turn B to select “BREAK”, then press A.
The measurement screen will ap-
pear.
6
Place the specimen measuring port to the specimen, and then press C (MEAS.
button).
The color difference from the target color, pass/fail judgment result based on the specified box tolerances
and direction of tint and brightness deviation will be displayed.
The color space to be used as the assessment axis differs depending on the selected color space.
When “L*C*h, E”, “L*C*h, CMC1:1”, “L*C*h, CMC2:1” or “L*C*h, CIE94” is selected, assess-
ment will be made based on the direction of tint and lightness deviation and color difference from the
target color in the L*C*h axis.
When a color space other than those given above is selected, assessment will be made based on the
direction of tint and lightness deviation and color difference from the target color in the L*a*b* axis.
M
emo
Assessment results differ as shown in the table below depending on the type of specified color difference toler-
ances and pass/fail judgment result.
Type
Specular
component
Displayed value
Pass/fail
judgment
Display content
Box
tolerance
I+E
Color difference
PASS Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
message to the right
E*
PASS Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
“FAIL” to the right
SCI or SCE
Color difference
PASS Measured value, with “PASS” to the right.
FAIL
Highlighted measured value, with highlighted
“FAIL” and message to the right
E*
PASS Measured value, with “PASS” to the right
FAIL
Highlighted measured value, with highlighted
“FAIL” to the right
Elliptical
tolerance
I+E
SCI or SCE
Color difference No judgment Measured value, with message to the right
E*
PASS Measured value, with “PASS” to the right
FAIL
Measured value, with highlighted “FAIL” to the
right