Konica Minolta CM-2600d Drums User Manual


 
E-36
Setting a Measurement Condition
Setting the UV
3
Turn B to select “UV SETTING”, then press A.
For CM-2600d
<Settings>
• UV100%: Measurement is performed with an illumination that contains all UV components of Xe flash
light source.
• UV0%: Measurement is performed with an illumination that contains no UV components of Xe flash light
source.
• UVADJUSTED: This item can be selected only if UV coefficients have been downloaded from Spec-
traMagic (Ver.3.2 or higher; excluded Ver.3.5).
For CM-2500d
Only “UV100%” can be selected.
Selecting Illuminant 1
4
Turn B to select the desired illuminant, then press A.
<Settings>
• D65: Standard illuminant D65; daylight, Color temperature: 6504K
• D50: Complementary illuminant D50; daylight, Color temperature: 5003K
C: Complementary illuminant C; daylight (Ultraviolet part is small in comparison with the daylight),
Color temperature: 6774K
• A: Standard illuminant A; Incandescent lamp, Color temperature: 2856K
• F2: Cool white (fluorescent lamp)
• F6: Cool white (fluorescent lamp)
• F7: Color rendering A daylight white (fluorescent lamp)
• F8: Color rendering AAA natural white (fluorescent lamp)
• F10: 3-band type natural white (fluorescent lamp)
• F11: 3-band type cool white (fluorescent lamp)
• F12: 3-band type warm white (fluorescent lamp)
• ---: None (Not selectable “ILLUMINANT 1”)