E-75
Displaying the Measurement Results
Measurement Results for “linked to each data.”
This instrument can set multiple target color data numbers and select one of those for displaying the re-
sults for the color difference. It can also easily switch the selected target color number using the naviga-
tion wheel. The “defined in COND.” target mode can use these functions to calculate multiple color
differences for measurement data, and make pass/fail judgments based on this.
However, the pass/fail judgment is based on a standard color difference setting for the selected target
color data number and the color difference. The differences between the target color data and the color
difference is not always standard.
The other target mode supported by this instrument, “linked to each data.”, can store the target color data
number with the selected measurement data and display the results using the associated target color data
number. In addition to absolute values, it can also be used for general color differences.
This section describes how the results are affected when using “linked to each data.” target mode and the
specified target color data has been changed or deleted. It also compares the difference to the results dis
-
played for “defined in COND.” target mode.
When the Color Difference Target Data No. has Been Changed
When “T” is selected in the measurement results screen, the target color screen will be changed to the
currently selected number to be used for next measurement. The navigation wheel can also be used in
this screen to recall a target color number that has been stored in the instrument’s memory. If “BREAK”
is selected in the target color data screen, you can revert to the originally selected target color number.
For details, see “Switching between the Measured Data screen and the <TARGET> screen” on page E-
49.
Select D, press A and then turn B to the
left or right, it can be switched the target color data
number.
In “linked to each data.” mode, the measured data is cal-
culated with the target data which is selected at the time
of measurement and the results are displayed.
Select “BREAK” and then press A, the dis-
play return to the measurement screen.
In “defined in COND.” mode, the measured data
is recalculated with the new selected target color
number’s data and the results are displayed.
Select and then press A, the <TARGET> screen
is displayed. It shows the target color data which number
is currently selected and is used for next measurement.