16
MODEL 3081 pH/ORP SECTION 3.0
WIRING
Table 3-4 Wiring Diagrams for Model 396P Sensors
Sensor Junction Box Preamplifier RTD Wire Function Wiring Diagram
396P-01-55 none in sensor Pt 100 Figure 3-13 Figure 3-14
396P-01-55 remote in sensor Pt 100 Figure 3-13 Figure 3-14
396P-02-50 none in transmitter 3K Balco** Figure 3-10 Figure 3-11
396P-02-50 remote in remote junction box 3K Balco** Figure 3-10 Figure 3-12
396P-02-54 none in transmitter Pt 100 Figure 3-10 Figure 3-11
396P-02-54 remote in remote junction box Pt 100 Figure 3-10 Figure 3-12
396P-02-55 none in transmitter Pt 100 Figure 3-10 Figure 3-11
396P-02-55 remote in remote junction box Pt 100 Figure 3-10 Figure 3-12
Table 3-5 Wiring Diagrams for Model 396 Sensor
Sensor Junction Box Preamplifier RTD Wire Function Wiring Diagram
396-50* none in transmitter 3K Balco** Figure 3-6 Figure 3-8
396-50* remote in remote junction box 3K Balco** Figure 3-6 Figure 3-9
396-54* none in transmitter Pt 100 Figure 3-6 Figure 3-8
396-54* remote in remote junction box Pt 100 Figure 3-6 Figure 3-9
396-54-62 none in transmitter Pt 100 Figure 3-7 Figure 3-8
396-54-62 remote in remote junction box Pt 100 Figure 3-7 Figure 3-9
Table 3-6 Wiring Diagrams for Model 389 Sensors
Sensor Junction Box Preamplifier RTD Wire Function Wiring Diagram
389-02-50* none in transmitter 3K Balco** Figure 3-6 Figure 3-8
389-02-50* remote in remote junction box 3K Balco** Figure 3-6 Figure 3-9
389-02-54* none in transmitter Pt 100 Figure 3-6 Figure 3-8
389-02-54* remote in remote junction box Pt 100 Figure 3-6 Figure 3-9
389-02-54-62 none in transmitter Pt 100 Figure 3-7 Figure 3-8
389-02-54-62 remote in remote junction box Pt 100 Figure 3-7 Figure 3-9
* Sensors have a BNC connector that the Model 3081 pH/ORP transmitter does not accept. Cut off the BNC and terminate
the coaxial cable as shown in Figure 3-23. Alternatively, use a BNC adapter (PN 9120531).
** Set the RTD jumper to the 3K position (see Section 2.2). Also, program the transmitter to recognize the 3K RTD (see
Section 8.5 for pH or 10.5 for ORP).