5.7 SECURITY
5.7.1 General. Use the programmable security code to protect program and calibration
settings from accidentally being changed. The transmitter is shipped with the security fea-
ture disabled. To program a security code, refer to Section 8.6, Display Units.
5.7.2 Entering the Security Code.
1. If calibration and program settings are protected with a security code, pressing PROG
or CAL on the infrared remote controller causes the Id screen to appear.
2. Use the editing keys to enter the security code. Press ENTER .
3. If the security code is correct, the first sub-menu appears. If the security code is incor-
rect, the process display reappears.
5.7.3 Retrieving a Lost Security Code.
1. If the security code has been forgotten, enter 555 at the Id prompt and press ENTER .
The transmitter will display the present code.
2. Press EXIT to return to the process display.
3. Press PROG or CAL . The Id screen appears.
4. Use the editing keys to enter the security code just shown; then press ENTER .
5. The first sub-menu under the selected menu will appear.
MODEL 3081 pH/ORP SECTION 5.0
OPERATION WITH REMOTE CONTROLLER
PROGRAM
I d
EXIT ENTER
000