75
10.1 GENERAL
This section describes how to do the following:
1. assign ORP values to the 4 and 20 mA outputs,
2. set the current generated by the transmitter during hold,
3. set the current generated by the transmitter when a fault is detected,
4. change sensor diagnostic limits,
5. change the units of the displayed variables,
6. program a security code,
7. simulate output currents for testing.
Factory default settings are given in Table 10-1. If default settings are acceptable, the transmitter is ready for cal-
ibration. See Section 9.0, Calibration of ORP Measurements. There is no way to automatically reset the transmit-
ter to factory defaults. Settings must be returned to default values one at a time. Figure 5-5 shows the menu tree.
10.2 ENTERING AND LEAVING THE PROGRAM MENU
Press PROG on the infrared remote controller (IRC) to enter the Program menu. To save new settings, press
ENTER . To leave the Program menu without saving new values, press EXIT . Pressing EXIT with a prompt
showing returns the display to the first prompt in the sub-menu. Pressing EXIT again returns the transmitter
to the process display.
If program settings are protected with a security code, pressing PROG or CAL will cause the Id screen to
appear. Key in the security code and press ENTER . The first sub-menu will appear. For more information,
see Section 5.7, Security.
A transmitter adjacent to the one being programmed may pick up signals from the IRC. To avoid accidentally
changing settings, use a different security code for each nearby transmitter. See Section 5.7, Security, and Section
10.6, Display Units, for details.
MODEL 3081 pH/ORP SECTION 10.0
PROGRAMMING FOR ORP MEASUREMENTS
SECTION 10.0
PROGRAMMING FOR ORP MEASUREMENTS
10.1 General
10.2 Entering and Leaving the Program Menu
10.3 Output Ranging
10.4 Diagnostic Parameters
10.5 Temperature Element
10.6 Display Units
10.7 Generating a Test Current