Fluke PM3370B Musical Instrument User Manual


 
4 - 72 HOW TO USE THE INSTRUMENT
AUTOSET USER PROGRAMMABLE
The AUTOSET function can also be programmed so that certain functions switch
to a predefined position after an AUTOSET.
USERPROG allows the user to customize the AUTOSET function for specific
applications.
Example: To program the AUTOSET function for dc coupling of the inputs, do
the following:
- Press the UTILITY key.
- Press the softkey labeled ’AUTOSET’.
- Select ’userprog’
- Press the ’VERT >’ softkey; select ’dc’, and press RETURN.
- Press the AUTOSET key.
The Probe Adjust signal is now displayed as dc coupled, instead of ac coupled as
would be the case after a STANDARD AUTOSET.
SETUPS
All settings of all controls can be stored in any of 10 memory locations.
This is a useful feature when performing routine tasks involving complex
setup procedures. Rather than changing all parameters to go from one
measurement to another, the oscilloscope can recall even the most complex
settings combinations with the touch of one button.
The SETUPS key is used to store and recall previously defined instrument
settings.
First clear all memory locations.
- Press the SETUPS key.
- Press the ’CLEAR&PROTECT’ softkey.
- Press the ’clear all’ softkey.
- Press the ’yes’ confirm softkey.
Example: Set the attenuator to 500 mV/div, and set the timebase to 500 µs/div.
- Position the trace in the bottom of the screen.
SETUPS