Fluke PM3370B Musical Instrument User Manual


 
FUNCTION REFERENCE 5 - 69
STANDARD SETUP / FRONT PANEL RESET
Description:
A factory-programmed set of default settings is available to put the instrument in
a defined state. The default settings (std) are reached in the menu under the
SETUPS menu key. Another method to perform a front panel reset is by pressing
the STATUS and TEXT OFF keys simultaneously.
The TRACK control is used to Select ’std’ and the memories s1 ... s10 (refer to
SETUPS). The settings saved in ’std’ become the actual settings when the ’recall’
softkey is pressed.
The most important ’std’ settings are:
- Analog mode.
- AUTOSET USERPROG is set to ’off’.
- Setting readout is switched on.
- Ground indicator is set to ’on’.
- Usertext is set to ’off’.
- Cursors are ’off’.
- Input is dc coupled and input impedance is 1 M.
- Input sensitivity is set to 50mV and VAR off.
- CH1 is on. CH2, CH3, CH4 and EXT TRIG are off.
- Vertical POS is set to center of screen.
- MAIN TB on and X-DEFL ’off’.
- MAIN TB in 1 ms and auto mode. VAR and 10x MAGNIFY are off.
- MAIN TB triggering in edge, level-pp on, positive slope, ac coupled and trigger
source CH1.
Note: Error messages appear in the CRT viewing area when incorrect
commands are given.
Key sequence:
Remote commands:
CPL: DS (Command for a front panel reset).
Refer to Chapter 6 for full details
SETUPS
TRACK
recall
ST6767
9303
std memory selection
softkey to recall ’std’ settings.