THE CPL PROTOCOL 6 - 27
Response 1:
Response 2:
n meas_type
1)
suffix_unit
1)
Note:
1 variable variable measurement 1
2 variable variable measurement 2
10 dV V delta Voltage between cursors
11 V1 V Volt cursor 1
12 V2 V Volt cursor 2
13 Vdc V Volt dc
20 dT s delta Time between cursors
2)
21 F Hz Frequency (1/dT)
30 dX V delta X
3)
40 dPh o Phase difference
51 T1-trg s Time between cursor 1 and trigger
52 T2-trg s Time between cursor 2 and trigger
60 FFT_freq Hz Frequency at cursor
4)
61 FFT_ampl dB/dBm/
dBµV/Vrms Amplitude at cursor
4)
Notes:
1)
The meas_type and suffix_unit of the measurement 1 and 2 values
depend on the selections, made with the MEASURE MENU via the
front-panel key MEASURE. Refer to the MEASURE MENU of
chapter 5 "FUNCTION REFERENCE".
2)
Delta T can only be obtained, if X-Deflection is off.
3)
Delta X can only be obtained, if X-Deflection is on.
4)
FFT only when MATH+ option is availlable.
acknowledge CR
meas_value
","
meas_type ",""," meas_value suffix_unit CR
Response 2: Condition:
type,value,unit No V or L parameter
value V parameter specified
value{,value} L parameter specified