FUNCTION REFERENCE 5 - 65
NO DTB, ACQ. TOO LONG Del’d TB only possible at acquisition
length of 512 sample points.
Functions ’ACQUISITION LENGTH’ and
’DEL’D TB’.
NO DTB IN ROLL MODE Del’d TB and Roll modes cannot be
combined.
Function ’DEL’D TB’ and ’TB MODE’.
NO ENVELOPE IN ROLL MODE Envelope and Roll modes cannot be
combined.
Function ’ENVELOPE’ and ’TB MODE’.
NO EVENT DELAY, DUAL SLOPE Event delay and dual slope cannot be
combined.
Function ’DELAY’.
NO EVENT DELAY IN TV MODE Event delay and TV trigger mode cannot
be combined.
Function ’DELAY’ and ’TV TRIGGER’.
NO IND WITH THIS COUPLING Trigger level indication not possible at dc
input coupling and ac or lf-rej trigger
coupling.
Function ’UTILITY SCREEN & SOUND.
NOT PART OF X VS Y REGISTER X-source only possible out of chosen
register.
Functions ’DISPLAY MENU’ and
’X-DEFLECTION’.
PLEASE FIRST SWITCH TO DSO Activated function only possible in digital
mode.
Function ’ANALOG/DIGITAL MODE’.
PLOT KEY = ABORT Indicates that the print/plot action can
only be aborted with PLOT key.
Function ’PRINTING AND PLOTTING’.
PRESS 2 SEC FOR AUTOCAL CAL key must be pressed for more then
two sec to start autocalibration.
Function ’CALIBRATION AUTOCAL’.
PROBE DETECTED, NO CHANGE The automatic probe detection overrules
manual selection when a probe with
indication ring is used.
Function ’PROBE UTILITIES’.