Fluke PM3370B Musical Instrument User Manual


 
6 - 42 THE CPL PROTOCOL
RECALL SETUP RS
Purpose:
To recall an internally stored setup from one of 10 setup registers. This setup must
have been stored in the oscilloscope manually or with the Save Setup (SS)
command.
The command performs the same as the frontpanel key SETUPS together with
the softkey RECALL.
Command:
<setup reg> represents the setup register number ranging from 1...10
Response:
Note: <acknowledge> will be sent after the internal setup has been completed.
Example:
100 PRINT #1,"RS";5 :’Recall setting 5
110 GOSUB 1000 :’Sync on acknowledge
"RS" pds CRsetup reg
acknowledge CR