FUNCTION REFERENCE 5 - 77
Results of the measurements MEAS1 and MEAS2 are displayed in the top left
corner of the screen. When you press the key sequence MEASURE >> CURSOR
LIMIT & STATIST >> STATIST on, the screen displays three values per MEAS
function. These values are measured over the total number of acquisitions for that
particular measurement and are updated instantly. The statistic measurement
values are as follows:
- absolute minimum value
- absolute maximum value
- mean value
Key sequence:
Remote commands:
CPL: QM (Command to query a measurement)
Refer to Chapter 6 for full details.
MEASURE MEAS1(2)
TRACK
volt
time
delay
ST6770
9312
on off
CURSOR
LIMIT&
STATIST
CURSOR
LIMIT
yes no
STATIST
on off
Toggle softkey to select time measurements in menu
MEAS1 or MEAS 2.
Control to choose the time measurement.
Control to select the signal on which the measurement
must be made.
Toggle softkey to switch the measurement on and off.
Toggle softkey to switch cursor limited measurements
on and off.
Toggle softkey to switch the statist function on and off.