THE CPL PROTOCOL 6 - 43
SAVE SETUP SS
Purpose:
To save the current oscilloscope setup in one of 10 internal setup registers. This
setup can be recalled manually or by sending the Recall Setup (RS) command.
The command performs the same as the frontpanel key SETUPS together with
the softkey SAVE. An execution error will be returned if the setup register is write
protected.
Command:
<setup reg> represents the setup register number, ranging from 1 ... 10
Response:
Example:
100 PRINT #1,"SS";3 :’Save setup in reg 3
110 INPUT#1,ACK : Sync on acknowledge
"SS" pds CRsetup reg
acknowledge CR