Fluke PM3370B Musical Instrument User Manual


 
THE CPL PROTOCOL 6 - 9
CALIBRATE CL
Purpose:
To start the internal Auto-Calibration procedure. This procedure optimizes the
input, trigger and time base circuitry of the oscilloscope. This calibration takes
approximately one minute and completion is signalled by the acknowledge.
The Calibrate (CL) command performs the same function as pressing the front
panel CAL button for more than 2 seconds.
Command:
Response:
Note: The <acknowledge> will be sent after the calibration has been fully
completed. A detailed error report may be queried for using the ST
command (only if acknowledge is not zero).
Example:
100 PRINT #1,"CL" :’Send command
110 GOSUB 1000 :’Sync on acknowledge
"CL" CR
acknowledge CR