5 - 58 FUNCTION REFERENCE
REMOTE CONTROL IEEE 488.2
Description:
An IEEE 488.2 Interface is available as an option. This interface can be used to
control oscilloscope functions by an external computer. All of the oscilloscope’s
current settings can be read by the computer.
The programming language is called SCPI (Standard Commands for
Programmable Instruments). SCPI is an IEEE standardized language designed
for remote control of programmable test and measuring equipment. General
information for SCPI and the instructions are located in a separate programming
manual. The 24-pole connector and its connections are shown in the figure below.
For correct functioning in a IEEE 488 environment, the oscilloscope’s device
address must be selected. This is done in the UTILITY REMOTE CONTRL menu
with the TRACK control. This menu is reached via the key sequence ’UTILITY >>
REMOTE CONTRL’. To change the IEEE settings, first select IEEE with the
’RS-232 IEEE’ softkey.
Operation of front key STATUS LOCAL
passes the control of the oscilloscope from
the interface (remote) to the front keys
(local).
Key sequence:
Key to switch from remote to local
SHIELD
ATN
SRQ
IFC
NDAC
NR
FD
DAV
DIO4
DIO2
EO1
DIO3
DIO1
GND
11
GND
9
GND
7
REN
DIO7
DIO5
LOGIC
GND
GND
10
GND
8
GND
6
DIO8
DIO6
ST6064
12
24
1
13
UTILITY
TRACK
REMOTE
SETUP
ST6075
9303
Control to select the IEEE device address.
STATUS
LOCAL