Teledyne T200H/M Drums User Manual


 
Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
108
4.11.4. ANALYZER ID
Each type of Teledyne API analyzer is configured with a default ID code. The default
ID code for all T200H/M analyzers is either 0 or 200. The ID number is only important
if more than one analyzer is connected to the same communications channel such as
when several analyzers are on the same Ethernet LAN (see Section 4.11.7); in a RS-232
m
ultidrop chain (see Section 4.11.9) or operating over a RS-485 network (see Section
4.11.6). If two analyzers of the same model type are used on one channel, the ID codes
of one or bot
h of the instruments needs to be changed so that they are unique to the
instruments. To edit the instrument’s ID code, press:
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNG
E
PASS CLK MORE EXIT
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X. MACHINE ID: 200 ID
0 2 0 0 ENTR EXIT
Toggle these buttons
to cycle through the
available character set:
0-9
ENTR button accepts the
new settings
EXIT key ignores the new
settings
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
The ID can be any 4 digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)
07270B DCN6512