244
7.5.4.3. AC Configuration –Heaters for Option Packages (JP6)
An O
2
sensor option includes AC heaters that maintain an optimum operating
temperature for key components of those options. Jumper set JP6 is used to connect the
heaters associated with those options to AC power. Since these heaters work with
either 110/155 VAC or 220/240 VAC, there is only one jumper configuration.
Table 7-5: Power Configuration for Optional AC Heaters (JP6)
JUMPER
COLOR
HEATER(S)
MODEL’S
USED ON
1
JUMPER
BETWEEN
PINS
FUNCTION
1 to 8
Common
IZS
1
Permeation Tube
Heater
100s, 200s
1
&
400s
2 to 7
Neutral to Load
3 to 10
Common
RED
O2 Sensor Heater 100s & 200s
4 to 9
Neutral to Load
1
IZS option not available on the T200H/M
6 5 4 3 2 1
IZS
(option not
available on the
T200H/M)
Permeation
Tube
Heater
O
2
Sensor
Heater
10 9
12 11 8 7
Figure 7-14: Typical Set Up of AC Heater Jumper Set (JP6)
07270B DCN6512