Teledyne T200H/M Drums User Manual


 
APPENDIX A-2: Setup Variables For Serial I/O T200H/M and 200EH/EM Menu Trees (05147H DCN6512)
A-14
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
200
9
1–500
9
CONC_RANGE3
1
Conc. 500,
100
4
,
20
9
1–20000,
1–10000
4
,
1–500
9
D/A concentration range 3 or
range for NO
2
.
CO2_RANGE
15
% 15 0.1–500 CO
2
concentration range.
O2_RANGE
14
% 100 0.1–500 O
2
concentration range.
50
3,4
,
40
5
RCELL_SET ºC
Warnings:
45–55
3,4
,
35–45
5
30–70
Reaction cell temperature set
point and warning limits.
50
4,6,8
,
40
5
MANIFOLD_SET
5
ºC
Warnings:
45–55
4,6,8
,
35–45
5
30–70
Manifold temperature set point
and warning limits.
CONV_TYPE — MOLY
3,5, 9
,
CONV
4
,
O3KL
6
NONE, MOLY,
CONV, O3KL
Converter type. “CONV” is mini-
hicon. Enclose value in double
quotes (") when setting from the
RS-232 interface. Changing this
variable changes CONV_SET
accordingly.
315,
200
6
CONV_SET ºC
Warnings:
305–325,
190–210
6
0–800
Converter temperature set point
and warning limits.
30 BOX_SET ºC
Warnings:
7–48
0–70
Nominal box temperature set
point and warning limits.
7
3,4
,
5
5
PMT_SET ºC
Warnings:
5–12
3,4
,
3–7
5
0–40
3,4
,
-10–40
5
PMT temperature warning limits.
Set point is not used.
500,
290
4
,
360
4+14
,
250
9
,
320
9+14
SFLOW_SET cc/m
Warnings:
350–600,
200–600
4,9
,
300–700
4+14,
9+14
0–1000,
100–1000
4, 9
Sample flow warning limits. Set
point is not used.
07270B DCN6512