Teledyne T200H/M Drums User Manual


 
APPENDIX A-3: Warnings and Test Measurements T200H/M and 200EH/EM Menu Trees (05147H DCN6512)
A-22
Name
1
Message Text Description
Warnings
WCONVTEMP CONV TEMP WARNING
Converter temperature outside of warning
limits specified by CONV_SET variable.
WPMTTEMP PMT TEMP WARNING
PMT temperature outside of warning limits
specified by PMT_SET variable.
WAUTOZERO
WPREREACT
11
AZERO WRN XXX.X MV
PRACT WRN XXX.X MV
11
Auto-zero reading above limit specified by
AZERO_LIMIT variable. Value shown in
message indicates auto-zero reading at
time warning was displayed.
WHVPS HVPS WARNING
High voltage power supply output outside
of warning limits specified by HVPS_SET
variable.
WDYNZERO CANNOT DYN ZERO
Contact closure zero calibration failed
while DYN_ZERO was set to ON.
WDYNSPAN CANNOT DYN SPAN
Contact closure span calibration failed
while DYN_SPAN was set to ON.
WREARBOARD REAR BOARD NOT DET
Rear board was not detected during
power up.
WRELAYBOARD RELAY BOARD WARN
Firmware is unable to communicate with
the relay board.
WFRONTPANEL FRONT PANEL WARN
Firmware is unable to communicate with
the front panel.
WANALOGCAL ANALOG CAL WARNING
The A/D or at least one D/A channel has
not been calibrated.
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.
2
Engineering firmware only.
3
Current instrument units.
4
Factory option.
5
O
2
option.
6
User-configurable D/A output option.
7
Optional.
8
CO
2
option.
9
Concentration alarm option.
10
M200EUP.
11
T200U, T200U_NOy, M200EU and M200EU_NO
y
.
12
T-Series External analog input option.
07270B DCN6512