Teledyne T200H/M Drums User Manual


 
Getting Started Teledyne API - Model T200H/T200M Operation Manual
68
STEP 2 – Activate O
2
sensor stability function
To change the stability test function from NO
x
concentration to the O
2
sensor output,
press:
SETUP X.X STABIL_GAS:O2
NO NO2 NOX O2 ENTR EXIT
Press ENTR to keep
changes, then press
EXIT 3 times to return
to SAMPLE menu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTREXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X 2) STABIL_GAS=NOX
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X STABIL_GAS:NOX
NO NO2 NOX O2 ENTR EXIT
Continue pressing NEXT until ...
Note Use the same procedure to reset the STB test function to NO
x
when the O
2
calibration procedure is complete.
07270B DCN6512