Teledyne API - Model T200H/T200M Operation Manual Calibration Procedures
189
STEP THREE
Activate NO
2
measurement stability function.
SETUP X.X STABIL_GAS:NO2
NO NO2 NOX O2 ENTR EXIT
Press ENTR first,
then press EXIT 3
times to return to
SAMPLE menu
SAMPLE RANGE = 50.000 PPM CO =X.XXX
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X ENTER PASSWORD:818
8 1 8 ENTR EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X 2) STABIL_GAS=NOX
<PREV NEXT> JUMP EDIT PRNT EXIT
SETUP X.X STABIL_GAS:NOX
NO NO2 NOX O2 ENTR EXIT
Continue pressing NEXT until ...
07270B DCN6512