Teledyne T200H/M Drums User Manual


 
APPENDIX A-3: Warnings and Test Measurements T200H/M and 200EH/EM Menu Trees (05147H DCN6512)
A-24
Test Measurement Name Message Text Description
Test measurements
NO2 NO2=0.0 PPB
3
NO
2
concentration for current range.
NO2_1
7
NO2_1=0.0 PPB
3
NO
2
concentration for range #1.
NO2_2
7
NO2_2=0.0 PPB
3
NO
2
concentration for range #2.
NOX NOX=396.5 PPB
3
NOX concentration for current range.
NOX_1
7
NOX_1=396.5 PPB
3
NOX concentration for range #1.
NOX_2
7
NOX_2=396.5 PPB
3
NOX concentration for range #2.
NO NO=396.5 PPB
3
NO concentration for current range.
NO_1
7
NO_1=396.5 PPB
3
NO concentration for range #1.
NO_2
7
NO_2=396.5 PPB
3
NO concentration for range #2.
CO2RANGE
8, not 6
CO2 RANGE=100.00 PCT D/A #4 range for CO
2
concentration.
CO2SLOPE
8
CO2 SLOPE=1.000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
8
CO2 OFFSET=0.000
CO
2
offset, computed during zero/span
calibration.
CO2
8
CO2=15.0 % CO
2
concentration.
O2RANGE
5, not 6
O2 RANGE=100.00 PCT D/A #4 range for O
2
concentration.
O2SLOPE
5
O2 SLOPE=1.000
O
2
slope computed during zero/span
calibration.
O2OFFSET
5
O2 OFFSET=0.00 %
O
2
offset computed during zero/span
calibration.
O2
5
O2=0.00 % O
2
concentration.
TESTCHAN
5,6,8
TEST=3627.1 MV
Value output to TEST_OUTPUT analog
output, selected with TEST_CHAN_ID
variable.
XIN1
12
AIN1=37.15 EU
External analog input 1 value in
engineering units.
XIN2
12
AIN2=37.15 EU
External analog input 2 value in
engineering units.
XIN3
12
AIN3=37.15 EU
External analog input 3 value in
engineering units.
XIN4
12
AIN4=37.15 EU
External analog input 4 value in
engineering units.
XIN5
12
AIN5=37.15 EU
External analog input 5 value in
engineering units.
XIN6
12
AIN6=37.15 EU
External analog input 6 value in
engineering units.
XIN7
12
AIN7=37.15 EU
External analog input 7 value in
engineering units.
XIN8
12
AIN8=37.15 EU
External analog input 8 value in
engineering units.
07270B DCN6512