Teledyne T200H/M Drums User Manual


 
T200H/M and 200EH/EM Menu Trees (05147H DCN6512) APPENDIX A-2: Setup Variables For Serial I/O
A-17
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
REMOTE_CAL_MODE
3
LOW LOW,
HIGH,
CO2
15
,
O2
14
Range to calibrate during remote
calibration. Enclose value in
double quotes (") when setting
from the RS-232 interface.
PASS_ENABLE — OFF ON, OFF
ON enables passwords; OFF
disables them.
STABIL_FREQ Seconds 10 1–300
Stability measurement sampling
frequency.
STABIL_SAMPLES Samples 25 2–40
Number of samples in
concentration stability reading.
650
3,5
,
550
4
,
600
9
HVPS_SET Volts
Warnings:
400–900
3,5
,
400–700
4
,
450–750
9
0–2000
High voltage power supply
warning limits. Set point is not
used.
6 RCELL_PRESS_SET In-Hg
Warnings:
0.5–15
0–100
Reaction cell pressure warning
limits. Set point is not used.
RCELL_CYCLE Seconds 10 0.5–30
Reaction cell temperature control
cycle period.
RCELL_PROP 1/ºC 1 0–10
Reaction cell PID temperature
control proportional coefficient.
RCELL_INTEG — 0.1 0–10
Reaction cell PID temperature
control integral coefficient.
RCELL_DERIV — 0 (disabled) 0–10
Reaction cell PID temperature
control derivative coefficient.
MANIFOLD_CYCLE
5
Seconds 5 0.5–30
Manifold temperature control
cycle period.
MANIFOLD_PROP
5
1/ºC 0.2 0–10
Manifold PID temperature control
proportional coefficient.
MANIFOLD_INTEG
5
0.1 0–10
Manifold PID temperature control
integral coefficient.
MANIFOLD_DERIV
5
0.5 0–10
Manifold PID temperature control
derivative coefficient.
IZS_CYCLE
3
Seconds 2 0.5–30
IZS temperature control cycle
period.
IZS_PROP
3
1/ºC 1 0–10
IZS temperature PID proportional
coefficient.
IZS_INTEG
3
0.03 0–10
IZS temperature PID integral
coefficient.
IZS_DERIV
3
0 0–10
IZS temperature PID derivative
coefficient.