Teledyne T200H/M Drums User Manual


 
Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
145
4.13.5.9. Setting Data Update Rate for an Analog Output
The data update rate for the T200H/M analog outputs can be adjusted to match the
requirements of the specific DAS parameter chosen for each channel. For instance, if
the parameter NXCNC1 (NO
x
concentration; Range 1) is chosen for channel A1 on an
instrument set for dual gas measurement mode, it would be meaningless to have an
update rate of less than 30 seconds, since the NO
x
-No measurement cycle takes that long
to complete. On the other hand, if the channel were set to output the PMTDET voltage
or the temperature of the moly converter, it might be useful to have output updated more
frequently.
07270B DCN6512