Teledyne T200H/M Drums User Manual


 
T200H/M and 200EH/EM Menu Trees (05147H DCN6512) APPENDIX A-3: Warnings and Test Measurements
A-23
Table A-3: Test Measurements
Test Measurement Name Message Text Description
Test measurements
NONOXCONC NO=396.5 NOX=396.5
3
Simultaneously displays NO and NOX
concentrations.
RANGE
not 6
RANGE=500.0 PPB
3
D/A range in single or auto-range modes.
RANGE1
not 6
RANGE1=500.0 PPB
3
D/A #1 range in independent range mode.
RANGE2
not 6
RANGE2=500.0 PPB
3
D/A #2 range in independent range mode.
RANGE3
not 6
RANGE3=500.0 PPB
3
D/A #3 range in independent range mode.
STABILITY NOX STB=0.0 PPB
3
O2 STB=0.0 PCT
5
CO2 STB=0.0 PCT
8
Concentration stability (standard deviation
based on setting of STABIL_FREQ and
STABIL_SAMPLES). Select gas with
STABIL_GAS variable.
RESPONSE
2
RSP=8.81(1.30) SEC
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
SAMPFLOW SAMP FLW=460 CC/M Sample flow rate.
OZONEFLOW OZONE FL=87 CC/M Ozone flow rate.
PMT PMT=800.0 MV Raw PMT reading.
NORMPMT NORM PMT=793.0 MV
PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
AUTOZERO AZERO=1.3 MV Auto-zero offset.
HVPS HVPS=650 V High voltage power supply output.
RCELLTEMP RCELL TEMP=50.8 C Reaction cell temperature.
BOXTEMP BOX TEMP=28.2 C Internal chassis temperature.
REMBOXTEMP
10
REM BOX TMP=30.1 C Remote chassis temperature.
PMTTEMP PMT TEMP=7.0 C PMT temperature.
MANIFOLDTEMP
4
MF TEMP=50.8 C Bypass or dilution manifold temperature.
CO2CELLTEMP
8
CO2 CELL TEMP=50.8 C CO
2
sensor cell temperature.
O2CELLTEMP
5
O2 CELL TEMP=50.8 C O
2
sensor cell temperature.
IZSTEMP IZS TEMP=50.8 C IZS temperature.
CONVTEMP MOLY TEMP=315.0 C
Converter temperature. Converter type is
MOLY, CONV, or O3KL.
SAMPRESTTEMP
10
SMP RST TMP=49.8 C Sample restrictor temperature.
RCELLPRESS RCEL=7.0 IN-HG-A Reaction cell pressure.
SAMPPRESS SAMP=29.9 IN-HG-A Sample pressure.
NOXSLOPE NOX SLOPE=1.000
NOX slope for current range, computed
during zero/span calibration.
NOXOFFSET NOX OFFS=0.0 MV
NOX offset for current range, computed
during zero/span calibration.
NOSLOPE NO SLOPE=1.000
NO slope for current range, computed
during zero/span calibration.
NOOFFSET NO OFFS=0.0 MV
NO offset for current range, computed
during zero/span calibration.
07270B DCN6512