4.15.2.6. Remote Access by Modem
The T200H/M can be connected to a modem for remote access. This requires a cable
between the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable
(available from Teledyne API with part number WR0000024).
Once the cable has been connected, check to make sure the DTE-DCE is in the correct
position. Also make sure the T200H/M COM port is set for a baud rate that is
compatible with the modem, which needs to operate with an 8-bit word length with one
stop bit.
The first step is to turn on the MODEM ENABLE communication mode (Mode 64).
Once this is completed, the appropriate setup command line for your modem can be
entered into the analyzer. The default setting for this feature is
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0
This string can be altered to match your modem’s initialization and can be up to 100
characters long.
Note If Hessen Protocol Mode is active for a com port, operation via a modem
is not available on that port.
To change this setting press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X COM1 MODEM INIT:AT Y &D &H
<SET SET> EDIT EXIT
SETUP X.X COM1 MODEM INIT:[A]T Y &D &H
<CH CH> INS DEL [A] ENTR EXIT
SETUP X.X COM1 MODE:0
SET> EDIT EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET SET> EDIT EXIT
ENT
R
accepts the
new string and returns
to the previous menu.
EXIT ignores the new
string and returns to
the previous menu.
The <C H and CH> buttons
move the [ ] cursor left and
right along the text string
IN
S
inserts a
character before
the cursor location.
Press the [?]
key repeatedly to cycle through the
available character set:
0-9
A-Z
space ’ ~ ! # $ % ^ & * ( ) - _ =
+[ ] { } < >\ | ; : , . / ?
DEL deletes a
character at the
cursor location.
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
Select which
COM Port is
tested
EXIT returns
to the
previous
menu
07270B DCN6512