4.13.7.2. Optic Test
The optic test function tests the response of the PMT sensor by turning on an LED
located in the cooling block of the PMT (Fig. 10-15). The analyzer uses the light
emitted from the LED to test its photo-electronic subsystem, including the PMT and the
current to voltage converter on the pre-amplifier board. To make sure that the analyzer
measures only the light coming from the LED, the analyzer should be supplied with zero
air. The optic test should produce a PMT signal of about 2000±1000 mV. To activate
the electrical test press the following touchscreen button sequence.
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG OPTIC A1:NXCNC1=100PPM NOX=XXX.X
<TST TST> EXIT
SETUP X.X ENTER DIAG PASS: 818
8 1 8 ENTR EXIT
Press NEX
T
until…
DIAG OPTIC TEST
PREV NEXT ENTR EXIT
Press TST until…
DIAG ELEC PMT = 2751 MV NOX=X.X
<TST TST> EXIT
While the optic test is
activated, PMT should be
2000 mV ± 1000 mV
Note
This is a coarse test for functionality and not an accurate calibration tool.
The resulting PMT signal can vary significantly over time and also
changes with low-level calibration.
07270B DCN6512