Teledyne T200H/M Drums User Manual


 
Teledyne API - Model T200H/T200M Operation Manual Calibration Procedures
203
To program the sample sequence shown above, follow this flow chart:
SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT MODE SET EXIT
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL CALZ CZLS SETUP
SETUP X.X PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE EXIT
SETUP X.X MODE: DISABLE
D
NEXT ENTR EXIT
SETUP X.X MODE: ZERO–HI
PREV NEXT ENTR EXIT
SETUP X.X SEQ 2) ZERO–HI, 1:00:00
PREV NEXT MODE SET EXIT
SETUP X.X STARTING DATE: 01–JAN–02
<SET SET> EDIT EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0 4 SEP 0 3 ENTR EXIT
Toggle to
set day,
month &
year: DD-
MON-Y
Y
SETUP X.X STARTING DATE: 04
SEP
03
<SET SET> EDIT EXIT
Default
value
is ON
SETUP C.4 STARTING DATE: 04
SEP
03
<SET SET> EDIT EXIT
Toggle to set
time: HH:MM.
This is a 24 hr
clock. PM
hours are 13
-
Toggle
numbers to
set
number of
days
between
procedures
(1
367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4 DELTA DAYS: 1
<SET SET> EDIT EXIT
SETUP C.4 DELTA DAYS: 1
0 0 2 ENTR EXIT
SETUP C.4 DELTA DAYS:2
<SET SET> EDIT EXIT
SETUP C.4 DELTA TIME00:00
<SET SET> EDIT EXIT
SETUP C.4 DELTA TIME: 00:00
0 0 :3 0 ENTR EXIT
SETUP C.4 DELTA TIME:00:30
<SET SET> EDIT EXIT
SETUP C.4 STARTING TIME:14:15
<SET SET> EDIT EXIT
EXIT returns
to the SETUP
Menu
SETUP C.4 DURATION:15.0 MINUTES
<SET SET> EDIT EXIT
SETUP C.4 DURATION 15.0MINUTES
3 0 .0 ENTR EXIT
SETUP C.4 DURATION:30.0 MINUTES
<SET SET> EDIT EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4 CALIBRATE: OFF
<SET SET> EDIT EXIT
SETUP C.4 CALIBRATE: OFF
ON ENTR EXIT
SETUP C.4 CALIBRATE: ON
<SET SET> EDIT EXIT
Toggle key
between
Off and
ON
SETUP X.X TIMER ENABLE: ON
SET> EDIT EXIT
SETUP C.4 STA
R
TING TIME:00:00
<SET SET> EDIT EXIT
Toggle NEXT button until ...
07270B DCN6512