Teledyne T200H/M Drums User Manual


 
T200H/M and 200EH/EM Menu Trees (05147H DCN6512) APPENDIX A-2: Setup Variables For Serial I/O
A-9
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
Medium Access Level Setup Variables (929 password)
LANGUAGE_SELECT — ENGL ENGL,
SECD,
EXTN
Selects the language to use for
the user interface. Enclose value
in double quotes (") when setting
from the RS-232 interface.
MAINT_TIMEOUT Hours 2 0.1–100
Time until automatically
switching out of software-
controlled maintenance mode.
LATCH_WARNINGS — ON ON, OFF
ON enables latching warning
messages; OFF disables latching
DAYLIGHTSAVING_ENABLE — ON ON, OFF
ON enables Daylight Saving
Time change; OFF disables
DST.
BXTEMP_TPC_GAIN — 0 0–10
Box temperature compensation
attenuation factor.
RCTEMP_TPC_GAIN — 0 0–10
Reaction cell temperature
compensation attenuation factor.
RCPRESS_TPC_GAIN — 1 0–10
Reaction cell pressure
compensation attenuation factor.
SPRESS_TPC_GAIN — 1 0–10
Sample pressure compensation
attenuation factor.
CE_CONC1A — 1 0-10000
Target CE concentration cal pt A
for range 1.
CONV_EFF1A — 1 0.8–1.2,
0.1–2
6, 19
Converter efficiency cal pt A for
range 1.
CE_CONC1B
19
1 0-10000
Target CE concentration cal pt B
for range 1.
CONV_EFF1B
19
— 1 0.1–2
Converter efficiency cal pt B for
range 1.
CE_OFFSET1
19
— 1 0.1–2
CE linearization Offset for range
1.
CE_SLOPE1
19
— 1 -10–10
CE linearization Slope for range
1.
CE_CONC2A — 1 0-10000
Target CE concentration cal pt A
for range 2.
CONV_EFF2A — 1 0.8–1.2,
0.1–2
6, 19
Converter efficiency cal pt A for
range 2.
CE_CONC2B
19
1 0-10000
Target CE concentration cal pt B
for range 2.
CONV_EFF2B
19
— 1 0.1–2
Converter efficiency cal pt B for
range 2.
CE_OFFSET2
19
— 1 0.1–2
CE linearization Offset for range
2.
CE_SLOPE2
19
— 1 -10–10
CE linearization Slope for range
2.
NEG_NO2_SUPPRESS — ON ON, OFF
ON suppresses negative NO
2
in
during switching mode;
OFF does not suppress negative
NO
2
readings