Teledyne T200H/M Drums User Manual


 
Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
113
Table 4-13: LAN/Internet Configuration Properties
PROPERTY DEFAULT STATE DESCRIPTION
DHCP STATUS
On Editable
This displays whether the DHCP is
turned ON or OFF.
INSTRUMENT
IP ADDRESS
Configured by
DHCP
EDIT key
disabled when
DHCP is ON
This string of four packets of 1 to 3
numbers each (e.g. 192.168.76.55.) is
the address of the analyzer itself.
GATEWAY IP
ADDRESS
Configured by
DHCP
EDIT key
disabled when
DHCP is ON
A string of numbers very similar to the
Instrument IP address (e.g.
192.168.76.1.)that is the address of
the computer used by your LAN to
access the Internet.
SUBNET MASK
Configured by
DHCP
EDIT key
disabled when
DHCP is ON
Also a string of four packets of 1 to 3
numbers each (e.g. 255.255.252.0)
that defines that identifies the LAN the
device is connected to.
All addressable devices and
computers on a LAN must have the
same subnet mask. Any transmissions
sent devices with different assumed to
be outside of the LAN and are routed
through gateway computer onto the
Internet.
TCP PORT
1
3000
Editable
This number defines the terminal
control port by which the instrument is
addressed by terminal emulation
software, such as Internet or Teledyne
API’ APICOM.
HOST NAME
[initially blank] Editable
The name by which your analyzer will
appear when addressed from other
computers on the LAN or via the
Internet. While the default setting for
all Teledyne API analyzers is the
model number, the host name may be
changed to fit customer needs.
1
Do not change the setting for this property unless instructed to by Teledyne API Technical
Support personnel.
Note It is recommended that you check these settings the first time you power
up your analyzer after it has been physically connected to the
LAN/Internet to confirm that the DHCP server has successfully
downloaded the appropriate information from you network. If the gateway
IP, instrument IP and subnet mask are all zeroes (e.g. “0.0.0.0”), the
DHCP was not successful. It may be necessary to manually configure the
analyzer’s Ethernet properties. Consult your network administrator.
07270B DCN6512