Teledyne T200H/M Drums User Manual


 
Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
141
The available gas concentration DAS parameters for output via the T200H/M analog
output channels are:
Table 4-23: Analog Output DAS Parameters Related to Gas Concentration Data
REPORTING
RANGE
PARAMETER
NAME
1
DESCRIPTION
NXCNC1
Concentration
NXSLP1
Slope
NXOFS1
Offset
NO
x
Range 1
(LOW)
NXZSC1
Concentration during calibration, prior to computing new slope and offset
NXCNC2
Concentration
NXSLP2
Slope
NXOFS2
Offset
NO
x
RANGE 2
(HIGH)
NXZSC2
Concentration during calibration, prior to computing new slope and offset
NOCNC1
Concentration
NOSLP1
Slope
NOOFS1
Offset
NO Range 1
(LOW)
NOZSC1
Concentration during calibration, prior to computing new slope and offset
NOCNC2
Concentration
NOSLP2
Slope
NOOFS2
Offset
NO RANGE 2
(HIGH)
NOZSC2
Concentration during calibration, prior to computing new slope and offset
N2CNC1
Concentration - Computed with data from NO
x
Range 1 and NO Range 1
NO
2
Range 1
2
(LOW)
N2ZSC1
Concentration during calibration, prior to computing new slope and offset
N2CNC2
Concentration - Computed with data from NO
x
Range 2 and NO Range 2
NO
2
RANGE 2
2
(HIGH)
N2ZSC2
Concentration during calibration, prior to computing new slope and offset
O2CONC
3
Concentration
O2OFST
3
Slope
O2SLPE
3
Offset
O
2
Range
3
O2ZSCN
3
Concentration during calibration, prior to computing new slope and offset
1
Parameters are not listed in the order they appear on the DAS list (see Table A-6 or Appendix A.5 for the proper order of the full list of
parameters)
2
Since NO
2
values are computed rather than measured directly, no separate slope or offset exist.
3
Only available on instruments with O
2
sensor options installed.
07270B DCN6512