Teledyne T200H/M Drums User Manual


 
Teledyne API - Model T200H/T200M Operation Manual Operating Instructions
149
4.13.6.1. Automatic Analog Output Calibration
To calibrate the outputs as a group with the AOUTS CALIBRATION command, press:
From the
AIO CONFIGURATION SUBMENU
(See Section 6.13.4.1)
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL is turned
ON
If any of the channels
have not been calibrated
ot if at least one channel
has AUTO-CAL turned
OFF, this message will
read NO.
DIAG AIO AUTO CALIBRATING DATA_OUT_1
DIAG AIO AUTO CALIBRATING DATA_OUT_2
DIAG AIO NOT AUTO CAL. DATA_OUT_3
DIAG AIO AUTO CALIBRATING DATA_OUT_4
This message
appears when
AUTO-CAL is
Turned OFF for
a channel
DIAG AIO AOUTS CALIBRATED: YES
SET> CAL EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
Note Manual calibration should be used for the 0.1V range or in cases where
the outputs must be closely matched to the characteristics of the
recording device.
07270B DCN6512