Teledyne T200H/M Drums User Manual


 
Troubleshooting & Repair Teledyne API - Model T200H/T200M Operation Manual
218
7.1.1. FAULT DIAGNOSIS WITH WARNING MESSAGES
The most common and/or serious instrument failures will result in a warning message
being displayed on the front panel. Table 4-3 lists warning messages, along with their
meaning and reco
mmended corrective action.
It should be d that if more than two or three warning messages occur at the same time, it
is often an indication that some fundamental analyzer sub-system (power supply, relay
board, motherboard) has failed rather than an indication of the specific failures
referenced by the warnings. In this case, a combined-error analysis needs to be
performed.
The analyzer will alert the user that a Warning message is active by flashing the red
FAULT LED, displaying the Warning message in the Param field along with the CLR
button (press to clear Warning message). The MSG button displays if there is more than
one warning in queue or if you are in the TEST menu and have not yet cleared the
message. The following display/touch screen examples provide an illustration of each:
The analyzer also issues an alert via the serial port(s).
07270B DCN6512