Teledyne T200H/M Drums User Manual


 
T200H/M and 200EH/EM Menu Trees (05147H DCN6512) APPENDIX A-2: Setup Variables For Serial I/O
A-13
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
NOX_SPAN2 Conc. 400,
80
4
,
20
11
,
16
9
0.01–9999.99
Target NO
X
concentration during
span calibration of range 2.
NO_TARG_ZERO2 Conc 0 -100–999.99
Target NO concentration during
zero calibration of range 2.
NO_SPAN2 Conc. 400,
80
4
,
20
11
,
16
9
0.01–9999.99
Target NO concentration during
span calibration of range 2.
NO2_SPAN2 Conc. 400,
80
4
,
20
11
,
16
9
0.01–9999.99
Target NO
2
concentration during
converter efficiency calibration of
range 2.
NOX_SLOPE2 PPM/mV 1 0.25–4 NO
X
slope for range 2.
NOX_OFFSET2 mV 0 -10–10 NO
X
offset for range 2.
NO_SLOPE2 PPM/mV 1 0.25–4 NO slope for range 2.
NO_OFFSET2 mV 0 -10–10 NO offset for range 2.
CO2_TARG_SPAN_CONC
15
% 12 0.01–100,
0.01–9999.99
16
Target CO
2
concentration during
span calibration.
CO2_SLOPE
15
1 0.5–5 CO
2
slope.
CO2_OFFSET
15
% 0 -10–10,
-100–100
16
CO
2
offset.
O2_TARG_SPAN_CONC
14
% 20.95 0.1–100
Target O
2
concentration during
span calibration.
O2_SLOPE
14
1 0.5–2 O
2
slope.
O2_OFFSET
14
% 0 -10–10 O
2
offset.
RANGE_MODE — SNGL SNGL,
IND,
AUTO,
REM
4,5
Range control mode. Enclose
value in double quotes (") when
setting from the RS-232
interface.
PHYS_RANGE1 PPM 2,
20
9
,
500
4
,
1
11
0.1–2500,
5–5000
9
,
5–10000
4
Low pre-amp range.
PHYS_RANGE2 PPM 22,
220
9
,
5500
4
,
100
11
0.1–2500,
5–5000
9
,
5–10000
4
High pre-amp range.
CONC_RANGE1 Conc. 500,
100
4
,
20
9
1–20000,
1–10000
4
,
1–500
9
D/A concentration range 1 or
range for NO
X
.
CONC_RANGE2
1
Conc. 500,
100
4
,
1–20000,
1–10000
4
,
D/A concentration range 2 or
range for NO.
07270B DCN6512