Teledyne T200H/M Drums User Manual


 
Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
174
To initialize the modem press:
Select which
COM Port is
tested
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
EXI
T
returns
to the
previous
menu
SETUP X.X COM1 MODEM INIT:AT Y
&D
&H
<SET SET> EDIT EXIT
S
ETUP X.X COM1 INITIALIZE MODEM
<SET SET> INIT EXIT
SETUP X.X COM1 MODE:0
SET> EDIT EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET SET> EDIT EXIT
EXIT returns to the
Communications Menu.
S
ETUP X.X INITIALIZING MODEM
<SET SET> INIT EXIT
4.15.2.7. COM Port Password Security
In order to provide security for remote access of the T200H/M, a LOGON feature can be
enabled to require a password before the instrument will accept commands. This is done
by turning on the SECURITY MODE (see Section 4.9). Once the SECUR
ITY
MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be
achieved with the LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause
subsequent logins to be disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
o LOGON SUCCESSFUL - Correct password given
o LOGON FAILED - Password not given or incorrect
o LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the T200H/M analyzer with SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable
RS232_PASS issued as follows:
V RS232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
07270B DCN6512