4.12.1. SETTING THE GAS MEASUREMENT MODE
In its standard operating mode the T200H/M measures NO, NO
2
and NO
x
. It can also
be set to measure only NO or only NOX. To select one of these three measurement
modes, press:
SETUP X.X 1 ) MEASURE_MODE=NOX-NO
PREV NEXT JUMP EDIT PRNT EXIT
EXIT ignores the new
setting.
ENTR accepts the
new setting.
SAMPLE ENTER SETUP PASS : 81
8
8 1 8 ENTR EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X MEASURE MODE: NO
X
-NO
PREV ENTR EXIT
SETUP X.X MEASURE MODE: NOX
PREV NEXT ENTR EXIT
SETUP X.X MEASURE MODE: NO
NEXT ENTR EXIT
NO
X
-NO mode is the
default mode for the
200EH/M
Press the PREV
and NEXT buttons
to move back and
forth between gas
modes
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15 minutes
NEXT JUMP EDIT PRNT EXIT
07270B DCN6512