Teledyne T200H/M Drums User Manual


 
Operating Instructions Teledyne API - Model T200H/T200M Operation Manual
86
4.7.2.2. Editing DAS Data Channels
DAS configuration is most conveniently done through the APICOM remote control
program. The following sequence of touchscreen button presses shows how to edit
using the front panel.
Edit Data Channel Menu
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL SETUP
SETUP X.X 0) CONC: ATIMER, 8, 800
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X DATA ACQUISITION
VIEW
EDIT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
EXIT will return to the
previous SAMPLE
display.
SETUP X.
X
ENTER DAS PASS: 818
8 1 8 ENTR EXIT
Moves the
display up &
down the list of
Data Channels
Inserts a new Data
Channel into the list
BEFORE the Channel
currently being displayed
Deletes The Data
Channel currently
being displayed
Exports the
configuration of all
data channels to
RS-232 interface.
Exits to the Main
Data Acquisition
Menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRNT EXIT
Moves the display
between the
PROPERTIES for this
data channel.
Reports the configuration of current
data channels to the RS-232 ports.
EXITS returns to
the previous
Menu
Allows to edit the channel name, see next key sequence.
When editing the data channels, the top line of the display indicates some of the
configuration parameters. For example, the display line:
0) CONC : ATIMER, 4, 800
Translates to the following configuration:
Channel No.: 0
NAME: CONC
TRIGGER EVENT: ATIMER
PARAMETERS: Four parameters are included in this channel
EVENT: This channel is set up to record 800 data points.
07270B DCN6512