Teledyne T200H/M Drums User Manual


 
Troubleshooting & Repair Teledyne API - Model T200H/T200M Operation Manual
254
Disconnect the heater cable coming from the relay board and measure the
resistance between any two of the three heater leads with a multi-meter. The
resistance between A and B should be about 1000 and that between A and C
should be the same as between B and C, about 500 each. If any of these
resistances is near zero or without continuity, the heater is broken.
Temperature reports zero or overload (near 500° C). This indicates a disconnected
or failing thermocouple or a failure of the thermocouple circuit.
First, check that the thermocouple is connected properly and the wire does not
show signs of a broken or kinked pathway. If it appears to be properly connected,
disconnect the yellow thermocouple plug (marked K) from the relay board and
measure the voltage (not resistance) between the two leads with a multi-meter
capable of measuring in the low mV range. The voltage should be about 12 mV
(ignore the sign) at 315° C and about 0 mV at room temperature.
Measure the continuity with an Ohm-meter. It should read close to zero . If the
thermocouple does not have continuity, it is broken. If it reads zero voltage at
elevated temperatures, it is broken. To test the thermocouple at room temperature,
heat up the converter can (e.g., with a heat gun) and see if the voltage across the
thermocouple leads changes. If the thermocouple is working properly, the
electronic circuit is broken. In both cases, consult the factory.
If the converter appears to have performance problems (conversion efficiency is outside
of allowed range of 96-102%), check the following:
Conversion efficiency setting in the CAL menu. If this value is different from 1.000,
this correction needs to be considered. Section 5.2.5 describes this parameter in
detail.
Accuracy of NO2 source (gas tank standard). NO2 gas standards are typically
certified to only ±2% and often change in concentrations over time. You should get
the standard re-certified every year. If you use GPT, check the accuracy of the
ozone source.
Age of the converter. The NO2 converter has a limited operating life and may need
to be replaced every ~3 years or when necessary (e.g., earlier if used with continu-
ously high NO2 concentrations). We estimate a lifetime of about 10000 ppm-hours
(a cumulative product of the NO2 concentration times the exposure time to that
concentration). However, this lifetime heavily depends on many factors such as
absolute concentration (temporary or permanent poisoning of the converter is
possible), sample flow rate and pressure inside the converter, converter tempera-
ture, duty cycle etc. This lifetime is only an estimated reference and not a
guaranteed lifetime.
In some cases with excessive sample moisture, the oxidized molybdenum metal
chips inside the converter cartridge may bake together over time and restrict air flow
through the converter, in which case it needs to be replaced. To avoid this problem,
we recommend the use of a sample gas conditioner (Section 5.10). Section 6.3.4
descri
bes how to replace the NO2 converter cartridge.
With no NO2 in the sample gas and a properly calibrated analyzer, the NO reading
is negative, while the NO2 reading remains around zero. The converter destroys
NO and needs to be replaced.
With no NO2 in the sample gas and a properly calibrated analyzer, the NOX reading
is significantly higher than the actual (gas standard) NO concentration. The
converter produces NO2 and needs to be replaced.
07270B DCN6512