Tektronix VX4101A Drums User Manual


 
SCPI Commands for the VX4101A
VX4101A MultiPaq Instrument User Manual
3–9
Command Response
ABOR
FETC:COUN? 153
FETC:COUN? 153
NOTE. After the ABORt command, the instrument will take no more measure-
ments.
CALibrate Subsystem
CALibrate:ROSCillator
:CLEar
:MANual <frequency>
N/A
Parameter Values
<frequency> Calibration Value Stored in non-volatile memory
Parameter Value
<frequency> 9 MHz Frequency 11 MHz
SOUR:ROSC[?]
SOUR:ROSC:VAL[?]
CALibrate:ROSCillator:CLEar
This command negates the calibration for the currently selected reference
oscillator. The reference oscillator is flagged as uncalibrated in nonvolatile
memory and any time the current source is reselected, an error message will be
generated noting this fact. An uncalibrated reference oscillator is assumed to be
exactly 10 MHz. This command would typically be used when a reference
oscillator source has been changed (such as changing a Slot 0 controller) and the
instrument has not yet been calibrated with the new source.
Command Syntax
Query Syntax
*RST Value
Limits
Related Commands
Command Description