Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Counter
VX4101A MultiPaq Instrument User Manual
3–107
The possible parameters for this command are as follows:
“TINTerval:DELay:TIMe”
“TINTerval:DELay:EVENTS”
SENSe[1|2|10|20]:FUNCtion
The SENSe:FUNCtion command selects a function and input channel without
changing most of the Counter setup. The input filtering, coupling, and imped-
ance are not changed by these commands. You must select the filtering, coupling
and impedance that makes sense for the input signal. The input attenuation,
offset, gain, and comparator hysteresis are not changed by these commands.
However, the comparator slopes and thresholds are changed to defaults by these
commands. If INPut:SETup:AUTO (autotrigger) is set to ON or ONCE, the
input attenuation, offset, gain, and comparator hysteresis may change when the
measurement is started. See the CONFigure or MEASure command descriptions
for detail of these commands.
The possible parameters for this command are as follows:
“TOTalize”
SENSe:TINTerval:DELay:EVENTs <event delay>
This command sets the number of events to use for the time interval with delay.
SENSe:TINTerval:DELay:TIME <time delay>
This command sets the delay time for time interval with delay by time function.
SENSe:APERture? [DEFault|MINimum|MAXimum]
This query moves the currently set aperture time, set by the most recent
SENSe:...:APERture, CONFigure or MEASure command, to the output buffer. If
one of the optional arguments is used, the DEFault, MINimum or MAXimum
value of aperture is moved to the output buffer instead.
SENSe:COUNT? [DEFault|MINimum|MAXimum]
This query moves the currently set <array size> count, as set by the most recent
SENSe:...:COUNt, CONFigure or MEASure command, to the output buffer. If
one of the optional arguments is used, the DEFault, MINimum or MAXimum
value of count is moved to the output buffer instead.
SENSe:EVENts? [DEFault|MINimum|MAXimum]
This query moves the number of events, as set by the most recent
SENSe:...:EVENts command, to the output buffer . If one of the optional
arguments is used, the DEFault, MINimum or MAXimum value of aperture is
moved to the output buffer instead.
Query Response