Tektronix VX4101A Drums User Manual


 
SCPI Commands for the Digital Input
3–130
VX4101A MultiPaq Instrument User Manual
Port Bits
1 1 through 8
2 9 though 16
3 17 through 24
4 25 through 32
If a bit_list is used, specific input bits will be enabled or disabled. Only bits
enabled as input bits can be read from and queried for its input. The bit list can
be comma separated or listed as (@x:y) for bits x through y.
MEASure returns the complete array (see READ?). To retrieve prematch data,
use a CONFigure or SENSe command followed by an INITiate command and
then a FETCh? query.
Command Response/Description
format hex Sets the return format to hexadecimal
sens:psel:digl enab Enables all bits as input
meas:digl? 2.3 #h00000000
Initiates a measurement with voltage threshold at 2.3 V
meas:digl:inv? 4.4 #hffffffff
Initiates a measurement with voltage threshold at 4.4 V and
inverts the bits.
READ? Subsystem
The Read subsystem initiates a measurement and returns the results.
N/A
READ?
N/A
N/A
Examples
Command Syntax
Query Syntax
*RST Value
Limits