Tektronix VX4101A Drums User Manual


 
SCPI Commands for the SurePatht Modules
3–268
VX4101A MultiPaq Instrument User Manual
will close channel 2 of section 1 of the VX4320.
A VX4330 can close only one channel in a group of joined sections that have
been specified in a [ROUTe:]CLOSe:MODe SCAN,<module_name>,<sec-
tion_list> command. If more than one channel in such a group of sections is
specified in a [ROUTe:]CLOSe <channel_list> command, the last channel
specified will be closed. For example, the commands
route:configure:join m2,(1:6)
route:close:mode scan,m2,(1:6)
join the commons of all six sections of the VX4330 Module and set the mode of
the [ROUTe]:CLOSE <channel_list> to scan mode for all six sections of the
VX4330.
The command
route:close (@m2(1!1,1!6))
will then result in channel 1 of section 6 being closed and all other channels on
the module being opened.
<section_list>
One or more <nr1> fields separated by comma (,) or colon (:) characters and
enclosed in left and right parentheses. A <section_list> is used to specify the
sections of a relay module to be acted upon by a [ROUTe:]CONFigure or
[ROUTe:]CLOSe:MODE command. The following are examples of valid
<section_list>s.
For commands directed to a VX4330 which has six scanner sections:
<section_list> Sections Specified
(1:6) Sections 1 through 6
(1,2,3) Sections 1, 2 and 3
(1:3,5:6) Sections 1 through 3 and 5 and 6
(1:3,5,6) Same as previous example
(3) Section 3
[ROUTe:]CLOSe <channel_list>
:DWELl <module_name>, <nrf>
:MODE <mode>,<module_name>,<section_list>
Command Syntax