The related commands for INPut subsystem are as follows:
MEASure
CONFigure
SENSe
INPut:IMPedance <impedance>
The INPut:IMPedance command defines the input impedance for DC voltage
measurements on the 30 mV, 300 mV, and 3 V ranges. The impedance may be
defined as either 10e6 or 10e9 W. Any value other than 10e6 or 10e9 is rounded
up to the nearer of these values. The 30 V and 300 V ranges are fixed at a
10 MW input impedance. If <Input Impedance> is not specified, 10e6 is
assumed. Specifying a value greater than 10e9 or less than zero generates an
error.
NOTE. If Autoranging is active for DC Voltage measurements, the impedance
value is automatically overwritten with 10.0e6 W when a DC Voltage measure-
ment is taken.
INPut:IMPedance?
Returns the currently defined impedance value.
Command Response
Input:Impedance 10e9 Define the input impedance as 10e9 W
INP:IMP? 1.00E+10<LF>
INSTrument Subsystem
The INSTrument commands enable you to control the operational state of the
DMM.
INSTrument:ABORt
INSTrument:RESet
N/A
N/A
Related Commands
Command Description
Query Response
Examples
Command Syntax
Query Syntax
*RST Value