Tektronix VX4101A Drums User Manual


 
SCPI Commands for the VX4101A
3–10
VX4101A MultiPaq Instrument User Manual
CALibrate:ROSCillator:MANual <frequency>
This command allows you to manually specify the calibrated frequency to be
used for the currently selected reference oscillator. The specified frequency value
is stored in non-volatile memory and overwrites the calibrated value stored
previously.
NOTE. To temporarily override the calibrated value without storing in non-vola-
tile memory, use the the SOUR:ROSC:VAL command.
CALibrate:ROSCillator:CLEar
Command Response
SOUR:ROSC
INT:SYST:ERR?
–313,“Calibration memory lost; VX4101A; Reference Oscillator
Uncalibrated”
SOUR:ROSC:VAL? 10000000.0000
INST:SEL COUNTER
CAL:ROSC 10.001E6
SOUR:ROSC:VAL? 9999999.0000
CAL:ROSC:CLE
SOUR:ROSC:VAL? 10000000.000
SOUR:ROSC:VAL? 10000000.0000
CALibrate:ROSCillator:MANual
Command Response
SOUR:ROSC INT
SYST:ERR –300 “Device specific error; VX4101A; Reference Oscillator
Not Calibrated”
CAL:ROSC:MAN 9.999999E6
SOUR:ROSC INT
SOUR:ROSC:VAL? 9999999.0000
SYST:ERR 0, “No error”